TitraLab
®
Add. reagent =
Titrant
Addition
Addition delay
The reagent used for addition is the same as the titrant used for the
titration.
Press ✓ and define an addition reagent different from the titrant in the
Parameters screen of the method.
Select Yes to carry out a reagent addition at the start of the titration or
a reagent calibration. This addition takes place before the titration
starts (predose included).
The reagent is added using a second burette controlled by the Titration
Manager. This second burette can be installed on the Titration
Manager or on an ABU52 connected to the Titration Manager,
"ABU1/ABU2", page
Edit an Addition method when you want to perform 2 or 3
reagent additions successively or simultaneously.
Refer to "Addition method - definition", page
Enter in:
Edit method > Parameters
Edit reagent > Calibration parameters
Parameter of an Addition method,
page
48.
For a 1-addition method or a multi-addition method with the
Simultaneous additions = Yes option selected, the
Addition delay is counted down at the end of the reagent
addition(s).
For a multi-addition method with the Simultaneous
additions = No option selected, an Addition delay is
to be set for each addition. An Addition delay is counted
down at the end of an addition and before the next addition is initiated.
While an addition method is running, you can carry on
regardless this delay by pressing key Del.
Enter in:
Edit method (Addition method).
Range available:
00:00 to 99:59 min:s
TitraLab 870 Reference Manual
43.
see "Addition method - definition",
see
48.
Page 47