TitraLab
®
Supervisor
mode
T°C minimum/
maximum value

Target titre

Temp. coef.

In SUPERVISOR mode, you can create, select, edit, delete sequence
of methods electrodes and reagents.
To select the Supervisor mode, a supervisor code is required.
Proceed as follows:
• Press Stop for 3 seconds in the Main window.
• Enter the Supervisor code.
• Press 5 Exit
• Select Return in mode = Supervisor.
Refer to "User's rights", page
If the temperature measured lies outside these limits a warning
message will be displayed.
Enter in:
Edit method > Results (for a measurement method using a tempera-
ture sensor).
Range available:
T°C minimum value: -9 °C to T°C maximum value.
T°C maximum value: T°C minimum value to 100 °C.
Approximate value of the reagent titre.
Enter in:
Reagent window > New reagent
Format:
5 characters
Refer to "Temp. correction None/Linear/Nat. water", page
TitraLab 870 Reference Manual
252.