Titrant ID

Titre Enter/
Calibrate
Tray missing
(SAC error)
The name of the titrant used by the method.
The titrant can be selected from the Users list of available reagents or
created from the preprogrammed list of reagents.
If the titrant is calibrated, the electrode(s) used for the calibra-
tion must be the same electrode(s) used in the method.
Enter in:
Edit method > Parameters menu
Enter
If you already know the reagent concentration and do not wish to per-
form a calibration.
Calibrate
If you wish to perform a calibration. The Titration Manager will prompt
you to enter the calibration parameters.
It is not necessary to calibrate an addition reagent. It is there-
fore recommended to select Titre = Enter so that the calibration
parameters are not displayed.
Enter in:
Edit reagent menu.
On starting a sequence, the sample changer cannot read the RFID
tag of the turntable. Check that a turntable is correctly mounted on the
sample changer.
Check the model of turntable used and refer to the User's Guide of the
sample changer, chapter 6 "Accessories".
In particular, note that a SAC850 turntable can be used on a SAC950
but a SAC950 turntable cannot be used on a SAC850.
Solve the problem then restart the sequence.