Result Id; Result Indicators - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Result ID

Result
indicators
Page 206
The name assigned by default to a method result number n is Rn (n=1
to 8). You can change the Rn name of each method result (16
alphanumerical characters maximum).
Enter in:
Edit method > Results
Edit electrode > Results
Edit reagent > Results
Refer to "Results menu", page
If you change a method result name (for example, from R1 to
«My result»), the equations taking this result into account
continue to use the original operand name (R1 in our
example).
Refer to "Equation formula", page
Appear with the result at the end of a run.
• OK: accepted by the instrument (result lies within the acceptance
limits).
• Low: alarm, rejected by the instrument (result lies below the
acceptance limit or belw the C
Direct ISE measurement method).
• High: alarm, rejected by the instrument (result lies above the
acceptance limit or above 10
ard addition ISE method).
• Time max: the measurement has been accepted at the end of the
Acceptance delay.
• QC: the user has bypassed a QC sample analysis demand.
• Man. stop: result has been obtained from a titration that was
stopped manually (Del key).
• 0IP: no inflection points found.
• IP>1: more than one inflection point has been found in the min/max
ordinate window,
see "Min. ordinate Max. ordinate", page
• Vstd>max: In an ISE Standard addition method with automatic
determination of the volumes, a Vstd>max note is added to the
result if the cumulate volumes of the standard additions is higher
than 50 % of the sample test quantity.
Refer to "ISE Standard addition method - programmation", page
143.
You can reject a result that was accepted by the instrument.
You can accept a result that was rejected by the instrument
(Supervisor users only).
Refer to "Result accepted (Yes/No)", page
TitraLab 870 Reference Manual
210.
blank concentration in the case of a
o
30
in the case of a Direct ISE or Stand-
122.
155.
205.

Advertisement

Table of Contents
loading

Table of Contents