TitraLab
®
Function
Fuses
Page 132
Select the electrode function relative to the electrode in use.
The possible electrode functions are:
• pH,
• ISE,
• mV (i = 0),
• mV (i > 0),
• T°C,
• Reference,
• Ground,
• Conductivity.
Refer to "Electrode type", page
For continued protection replace the fuse with one of a high
interrupting capacity, same type and rating:
2 x fuses, slow blow, 1.0 A (5 x 20 mm), part no. 450-020.
To replace the fuses:
1. Switch off the instrument
2. Disconnect line cord
TitraLab 870 Reference Manual
118.
3. Remove the fuse holder
hook
4. Replace the used fuses with
5. Put the cap back in place
Figure 24: Fuse replacement
ones of the same type and rating