Function; Fuses - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Function

Fuses

Page 132
Select the electrode function relative to the electrode in use.
The possible electrode functions are:
• pH,
• ISE,
• mV (i = 0),
• mV (i > 0),
• T°C,
• Reference,
• Ground,
• Conductivity.
Refer to "Electrode type", page
For continued protection replace the fuse with one of a high
interrupting capacity, same type and rating:
2 x fuses, slow blow, 1.0 A (5 x 20 mm), part no. 450-020.
To replace the fuses:
1. Switch off the instrument
2. Disconnect line cord
TitraLab 870 Reference Manual
118.
3. Remove the fuse holder
hook
4. Replace the used fuses with
5. Put the cap back in place
Figure 24: Fuse replacement
ones of the same type and rating