Number Of Additions; Number Of Buffers; Number Of Cycles; Number Of Decimals - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Number of
additions
Number of
buffers
Number of
cycles
Number of
decimals
Set the number of additions to be performed automatically by an
Addition method or an ISE standard addition method.
Refer to "Addition method - definition", page
Refer to "ISE Standard addition method - definition", page
Enter in:
Edit method menu (Addition method)
Edit method > Parameters (ISE standard addition method)
Range available:
1 to 3 (Addition method)
1 to 9 (ISE standard addition method)
Available if Calibration request = Yes
Number of pH standards to be used for the calibration. Work with at
least two standards to calculate the electrode sensitivity. If one
standard is only used the zero pH is calculated.
Enter in:
Edit electrode menu
Range available:
1 to 5
Refer to "pH buffer", page
Available if Calibration request = Yes
Number of times the calibration is to be repeated, i.e. the number of
beakers to be prepared for each pH, conductivity or ISE standard.
Enter in:
Edit electrode menu
Range available:
1 to 9
Number of decimals (0 to 3) to be displayed and printed for the result.
Enter in:
Edit method > Results (for a pH measurement method)
TitraLab 870 Reference Manual
48.
169.
142.
Page 163