TitraLab
®

Expiry date

Expiry date
elapsed

Fill burette

Final dil.
amount
Page 130
The expiry date of a reagent is to be entered when installing or
replacing a reagent. The expiry date is given on the reagent bottle.
The reagent cannot be used once this date is exceed. The Titration
Manager prompts you to replace the reagent bottle.
Format used:
dd:mmm:yyyy
Use the UP and DOWN keys to select the month.
The reagent expiry date has elapsed. Replace the reagent.
The expiry date is entered while installing the reagent.
1. Enter the Burette functions menu:
2. Select the burette in the Address field,
50.
3. Press 1 Fill. The burette is filled at maximum speed using the
installed reagent. Approximate filling time: 25 s.
Amount of sample after dilution,
Enter in:
Edit method > Sample
Range available:
0.001 to 100000 ml
TitraLab 870 Reference Manual
see "Address", page
see "Aliquot", page
51.