Detailed; Detection Limit (C0); Dilution (Yes/No) - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Detailed

Detection limit
(C
)
0
Dilution (Yes/
No)
This parameter sets level of details of report printouts.
Detailed = Low
• The header only comprises the analysis name, time and date and
the instrument serial number. These data are printed on the same
line.
• Reagent or electrode calibration method: results are printed.
Detailed = Medium
This is the printout level selected by default.
• The header comprises the analysis name, time and date, the
instrument serial number and the laboratory coordinates.
• Reagent or electrode calibration method: results are printed.
Detailed = High
This is the printout level selected by default.
• The header comprises the analysis name, time and date, the
instrument serial number and the laboratory coordinates.
• Electrode data, electrode serial number, electrode calibration data
and results, burette serial number, reagent data and reagent
calibration data and results are printed.
• The buffer or standard data (name and batch number, potential
value) are printed.
• End point/Inflection point coordinates (potential, volume) are
printed (except for a reagent calibration).
Enter in:
Edit method > Printouts
Edit reagent > Printouts
Edit electrode > Printouts
Refer to "Direct ISE measurement method - definition", page
Select Dilution = Yes when you are analysing samples which
have been diluted. The Final dil. amount and Aliquot
are to be entered next.
Enter in:
Edit method > Sample menu
TitraLab 870 Reference Manual
95.
Page 93