Reagent Calibration Not Required; Reagent Calibration Parameters - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Reagent
calibration not
required
Reagent
calibration
parameters
Message appears at the start of a sequence, when a method
sequence had been programmed with a reagent calibration.
The reagent has been reprogrammed without a calibration.
Go to Sequence/Sample stack, Edit sequence screen and remove the
reagent calibration from the sequence.
This menu contains the general parameters concerning the reagent
calibration method (identification of the electrodes used in particular).
Titration parameters necessary to run the calibration are also pro-
grammed.
Do not forget to select the electrode and temperature sensor
created in this menu!
To access:
1. From the Reagent window, press 4.
2. Select the reagent to be edited.
3. Press 2 Edit reagent and check that the Titre = Cali-
brate option has been selected.
4. Edit the calibration general parameters.
5. Use the LEFT/RIGHT arrow keys to move to the last Edit reagent
display.
6. Press 1 Calibration parameters.
TitraLab 870 Reference Manual
Page 191