Cable capacity

Page 70
A cable of a given length has a given capacity. When the measured
conductance is low (below 4 µS), the cable capacity is not
negligible and must be taken into account.
Enter the cable capacitance when:
• measuring low conductances (below 4 µS),
• the cable capacity of the conductivity cell is greater than 350 pF.
The cable capacity is normally specified by the manufacturer. Cable
capacities of a few Radiometer Analytical conductivity cells are given
below :
Conductivity cell
CDC511T
CDC861T
CDC565
CDC749
CDC267-9 with cable A94L136
CDC267-9 with cable A94L336
CDC241-9 with cable A94L136
CDC241-9 with cable A94L336
XE100 with cable A94L136
XE100 with cable A94L336
Figure 13: Cable capacities of Radiometer Analytical conductivity cells
If you create a conductivity cell from the Catalogue list, the
cable capacity is automatically assigned to the
conductivity cell created (and cannot be modified).
Enter in:
When creating an electrode with the Conductivity function and the
option From = Other.
Refer to "Create electrode", page
Available limits:
0 to 1999 pF
TitraLab 870 Reference Manual
Cable capacity (pF)
500
500
440
170
70
200
70
200
70
200
87.