Start Timer; Static Rinses; Static Rinse Time; Statistics - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Start timer

Static rinses

Static rinse time

Statistics

Std add volume
Automatic/
Programmed/
This is the time delay the Titration Manager will respect before starting
the titration. If a predose is programmed, this delay will be counted
down at the end of the predose. The Start timer delay will allow the
solution to stabilise before performing the first measurement.
While an addition method is running, you can carry on
regardless this delay by pressing key Del.
Enter in:
Edit method > Parameters menu
Edit reagent > Calibration parameters menu
Range available:
00:00 to 99:59 min:s
Refer to "Number of static rinses", page
Refer to "Rinse time", page
When several tests are performed on the same sample, the mean and
standard deviations are calculated from the accepted tests.
Automatic
The user enters the potential jump due to all additions (from 5 to
100 mV) and the number of additions (1 to 9) to be performed.
The TIM870 determines automatically the potential jump due to one
addition then calculates the volume to be added for each addition.
Programmed
The user enters the volume of one addition (from0.01 to 999 ml) and
the number of additions (1 to 9) to be performed.
Select in:
Edit method > Parameters (for an ISE standard addition method)
Refer to "ISE Standard addition method - definition", page
Refer to "ISE Standard addition method - programmation", page
TitraLab 870 Reference Manual
166.
211.
142.
143.
Page 239