Results By Difference/Cumulate; Results Factor (Yes/No) - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Results by
difference/
cumulate
Results factor
(Yes/No)
Results by difference (ml)
Volumes between two End points or Inflection points are used for
result calculation.
Results by cumulate (ml)
Total volume delivered up to the End point or Inflection point is used
for result calculation.
This setting has no influence on the result calculation for the
first equivalence point detected.
Enter in:
Edit method > Results (End point, IP monotonic, IP dynamic or IP
continuous methods)
This multiplying factor is applied to all Rn (n = 1 to 8) results of the
method. Equation results En (n = 1 or 2) are not processed.
Access:
Edit method > Results
Proceed as follows to multiply all your method results by a factor:
• In the method Results menu, select Results factor =
Yes.
• Run the method:
There is no results factor for a reagent calibration method.
TitraLab 870 Reference Manual
Enter the results factor (from 0.001 to 10 000)
then press 1 to start the method.