Stoichiometric
coefficients
Stop after last
IP

Stop analysis

Page 242
Refer to "Reaction X Exc. + Y Titr", page
Refer to "Reaction X Smp + Y Titr", page
Refer to "Reaction X Smp + Y Exc", page
Refer to "Reaction X Std + Y Titr", page
Stops a Continuous IP, Monotonic IP or Dynamic IP method once the
last IP has been found.
The last IP is set using the Number of IP parameter.
Enter in:
Edit method > Parameters
Edit reagent > Calibration parameters
The Stop key enables you to stop a test and display the following
screen:
The Del key enables to stop an analysis and calculate the results if
the instrument acquired sufficient points to carry out these
calculations (for a titration method). The result of an interrupted test is
saved with a "Man. stop" note.
When you end the analysis on a sample and several tests
have been accepted, a mean and standard deviation is
calculated for each result and stored in the archives.
When you end the analysis and have not performed the
number of tests required for one sample, you have the choice
between starting a new test on that sample (key 1) or ending
the analysis (key 2).
TitraLab 870 Reference Manual
185.
186.
187.
188.
Press 1 to restart a titration in the same test
portion. The titrant volume dispensed before
titration stops is taken into account for the cal-
culations.
Press 2 to start a new test in the same, or
different sample or end the analysis.
Press 3 to end the measurements. The active
test is not saved