Minimum Value - Maximum Value - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Minimum value -
Maximum value
Page 158
If the acceptance criteria option has been set to Yes, enter the
acceptance range for the result. If the result lies outside these limits, a
"High" or "Low" warning message appears and the result is rejected
by the instrument.
Enter in:
Edit method > Results menu (Acceptance criteria = Yes)
Edit reagent > Results menu (Acceptance criteria = Yes)
Edit method > QC data menu
Range available:
Minimum value = 0.0 to Maximum value
Maximum value = Minimum value to 10
Unit:
Sample method: Result unit
Reagent calibration: depends on the Reagent unit selected
(eq/l if N, meq/l if mN, mol/l if M, mmol/l if mM).
In a Direct or Standard addition ISE measurement method,
the instrument rejects all results that is higher than 10
("High" is displayed) even if you do not enter acceptation
limits for the result.
In the same way, in a Direct ISE measurement method, the
instrument rejects all results that is lower than the C
concentration ("Low" is displayed) .
Select Acceptance criteria = No in the Results menu of the
method if you do not want to enter acceptance limits for the
result,
see "Acceptance criteria", page
TitraLab 870 Reference Manual
10
44.
30
0