Printouts setup

Printouts title

Perform the following, before printing:
1. Connect the printer to the Titration Manager.
2. Enter the Setup menu (Stop, 3 seconds from the Main window).
3. Press 1 Configuration.
4. Select Printer = 80 columns.
5. Select the printout format (Listing or Page by Page).
Refer to "Format (printouts)", page
6. Press Esc then 3 to customise the printouts, e.g. enter
workstation name.
7. In the Edit method > Printouts, Edit reagent > Printouts and Edit
electrode > Printouts menus define the contents of the printouts.
The Detailed = High/Medium/Low option set the amount of
information that will be printed automaticaly at the end of each
test.
Refer to "Detailed", page
8. In the Edit method > Printouts menu of an ISE standard addition
method, select whether you want to print or not the GRAN =
f(volume) curve. This curve will be printed automaticaly at the end
of each test.
Refer to "GRAN = f(Volume) curve", page
9. In the Edit electrode > Printouts menu of an ISE electrode
calibration method, select whether you want to print or not the
calibration curve (E = f(pC = -log C)). This curve will be printed
automaticaly at the end of each calibration cycle.
Refer to "Calibration curve of an ISE electrode", page
Title of the report printout (1 to 23 characters).
Enter in:
Edit method > Printouts
Edit reagent > Printouts
Edit electrode > Printouts
TitraLab 870 Reference Manual