Standard (Conductivity Standard) - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Standard
(conductivity
standard)
Selection of the conductivity standard used to calibrate the
conductivity cell. 8 standards are available:
Conductivity
standard
1 D KCl
0.1 D KCl
0.01 D KCl
0.1 M KCl
0.01 M KCl
0.001 M KCl
0.05 % NaCl
25 µS/cm NaCl
Figure 42: Conductivity standards available in the TIM870
The TIM870 determines the cell constant from the conductivity values
of the standard which are saved versus temperature.
Access:
1. From the Electrode window, press 4.
2. Select the Conductivity type electrode to be edited.
3. Press 2 Edit electrode and check that the
Titre = Calibrate option has been selected.
4. Use the LEFT/RIGHT arrow keys to move to the last display.
5. Press 2 Standard line.
6. At the ID line, select the standard used for the calibration of the
conductivity cell.
TitraLab 870 Reference Manual
Temperature
Radiometer Analytical
range
0 to 27 °C
S51M001 (500 ml)
0 to 50 °C
S51M002 (500 ml)
0 to 50 °C
S51M003 (500 ml)
0 to 36 °C
C20C250 (500 ml)
0 to 34 °C
C20C270 (500 ml)
0 to 30 °C
C20C280 (500 ml)
0 to 99.9 °C
S51M004 (500 ml)
0 to 100 °C
S51M013 (250 ml)
part no.
Page 235