Turntable Blocked (Sac Error); Type Of Method; User Id (Yes/No); User List - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Turntable
blocked (SAC
error)

Type of method

User ID (Yes/No)

User list

The turntable has been blocked or forced and or cannot function
properly.
Remove the obstruction and press Resume analysis (key 1) to
continue the sequence from the point it stopped.
Warning!
Do not change the turntable before restarting the sequence from the
beaker it stopped (key 1) or before restarting the sequence from the
next beaker (key 2). The sample changer identifies a turntable only
when a new sequence is initialized (equivalent to a keystroke on 3
End of sequence followed by 1 Run sequence).
The type of method is selected in the Add method menu while editing
a sequence. The following types are available:
• Sample only methods,
• Electrode calibration methods,
• Reagent calibration methods,
• Blank sample methods (Yes has been selected for Blank in the Edit
method menu),
• QC sample methods (Yes has been selected for QC sample in the
Edit method menu).
If specified, User ID will be asked for each time an analysis is run.
Enter in:
Setup menu > Configuration
List of user defined methods, electrodes and reagents. The list is ini-
tialised to the preprogrammed methods, electrodes and reagents.
TitraLab 870 Reference Manual
Page 251