Result unit

Unit used for the result.
Enter in:
Edit method > Results
Range available:
If Sample unit = ml or µl: mol, mmol, eq/l, meq/l, mol/l, mmol/l, g/l,
mg/l, mg/ml or µg/ml
If Sample unit = g or mg: mol, mmol, eq/kg, meq/kg, mol/kg, mmol/kg,
g/kg, mg/kg, mg/g, µg/g, % (w/w) or ppm (w/w).
The result can be expressed as a mass (g, µg, mg) or a volume (µl or
ml).
In a Direct ISE or a Standard addition ISE method, the
available units are:
eq/l, meq/l, mol/l, mmol/l, g/l, mg/l, mg/ml, µg/ml, % (w/v) or
ppm (w/v).
Warning!
In a Direct or a Standard addition ISE method, the ppm and
% units are expressed in weight/volume.
Nevertheless, if the standard (or addition) solution has been
prepared in % or ppm (weight/weight), the % or ppm result
will be in weight/weight.
You can also use the Results factor to convert in % or ppm
(weight/weight) a result obtained in % or ppm
(weight/volume).
Refer to "Results factor (Yes/No)", page
TitraLab 870 Reference Manual
209.