Create electrode

1. Enter the Electrode window.
2. Press 4 then 1.
3. Select the electrode function,
4. Press ✓ in the ID field.
5. Select From = Other.
The option From = Catalogue allows you to create an
electrode from a list of Radiometer Analytical electrodes.
6. Enter the electrode name (up to 16 alphanumeric characters).
7. In the Confirm creation screen:
• For pH, mV or ISE electrodes only; select the electrode type.
Refer to "Electrode type", page
• For combined pH or single pH electrodes; enter the internal pH
(pH int) of the electrode.
Refer to "pH int", page
• For combined pH, Metal/Redox or ISE electrodes or for a
Conductivity electrode; select if the electrode has a built-in
temperature sensor or not.
• For reference electrodes or combined pH, Metal/Redox, ISE
electrodes; enter the potential of the reference (in mV) versus the
Standard Hydrogen Electrode.
Refer to "Potential versus SHE", page
• For ISE electrodes only; enter the ion valency and molar weight
(in g/mol).
Refer to "Valency", page
• For conductivity cells only; enter the cable resistance (in Ω ) and
capacitance (in pF).
Refer to "Cable resistance", page
Refer to "Cable capacity", page
8. Press 1 to create the electrode. The Edit electrode menu is
displayed. Enter the electrode definition parameters.
TitraLab 870 Reference Manual
see "Electrode type", page
118.
170.
172.
253.
71.
70.
118.