TitraLab
®

Measurement

Measurement type for the method.
Enter in:
Edit method menu.
Edit reagent menu if Titre = Calibrate.
Range available:
pH measurement (pH), zero-current potential measurement (mV),
imposed current potential measurement (mV(i>0)).
For Measurement methods (Mode = Measurement :
page
159), 3 other options are available : Direct ISE, ISE Std Add and
Conductivity.
If you select mV(i>0), connect the double platinum electrode to
the Pt-Pt socket on the rear panel. One of the electrode's poles
is connected to the ground, so it is necessary to select TIM
ext. cell Gnd = No in the Setup window.
If you select Conductivity, connect the conductivity cell to the
EC socket on the rear panel. One of the EC socket pin is
connected to the ground, so it is necessary to select TIM
ext. cell Gnd = No in the Setup window.
TitraLab 870 Reference Manual
see "Mode",
Page 151