TitraLab
®

Valency

Viewing global
variables

Volume

Vstd>max
(result
indicator)
This parameter is available when creating an ISE electrode with the
option From = Other.
Refer to "Create electrode", page
Select the valency of the ion under study.
For example: Valency = -1 for a fluoride (F
Range available:
+1, +2, -1 or -2.
1. Enter the Main window:
2. Press 5 GLP-Archives,
3. Press 2 Global variable
4. Scroll the G1 to G10 global variables using the arrow keys.
In an ISE standard addition method with programmed volumes, the
user enters the volume of each addition being performed.
To avoid modifying the ionic strength of the solution by dilution effect,
it is recommended to add small amounts of standard. The addition
cumulate volumes should not exceed 10% of the sample test quantity.
Example:
5 additions programmed with a sample amount of 1 ml.
Enter a volume for one addition equal to or less than 20 µl and use a
1 ml burette to perform the 5 additions.
Enter in:
Edit method > Parametrers
(ISE standard addition method with Std add volume =
Programmed).
Refer to "ISE Standard addition method - definition", page
Refer to "ISE Standard addition method - programmation", page
Refer to "Result indicators", page
TitraLab 870 Reference Manual
87.
-
) ion selective electrode.
206.
142.
143.
Page 253