Back Titration No/Manual/Automatic - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Back titration
No/Manual/
Automatic
Back titration = No
This is a direct titration.
Back titration = Manual
This is a back titration where the excess reagent is added using a
separate burette or a pipette not controlled by the Titration Manager.
This reagent is defined using its titre and volume in the Parameters
screen. This reagent is not selected in the Reagent library.
Back titration = Automatic
This is a back titration where the excess reagent is added using a
second burette controlled by the Titration Manager. This second
burette can be installed on the Titration Manager or on an ABU52
connected to the Titration Manager,
The excess reagent is created or selected from the User list.
Enter in:
Edit method > Parameters menu
TitraLab 870 Reference Manual
see "ABU1/ABU2", page
43.
Page 57