Edit Sequence Menu - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Edit sequence
menu
1. Perform steps 1 to 6 of the "Programming sequence" procedure.
Refer to "Programming sequence", page
2. In the Sequence/Sample stack window (access: press 2
Sequence/Sample stack from the main window),
press 3 Edit sequence.
Refer to "Add method menu", page
Refer to "Insert method menu", page
3. At the line Number of samples, enter the number of times a
method must be repeated within the sequence.
At the line Beakers: [F;L], the instrument displays the positions F
and L occupied by the first and last beakers In the sequence.
4. If a SAC850 or SAC950 is used and declared in the Configuation
menu, you can also choose a sample preparation number used
for the selected method,
The last step of the sequence programmation consists of editing the
sample stack,
see "Sample stack", page
TitraLab 870 Reference Manual
181.
Select the methods to be included in
the sequence. Use the Add method
(key 1)/ Insert method (key 2)
procedures.
In the title bar, "x/y" (1/1) indicates the
position "x" occupied by the method in
the sequence and "y" the total number
of methods listed in the sequence.
The ID and type of the selected
method cannot be modified at this
level. They are defined in the Add
method or Insert method
menu.
46.
139.
see "Sample preparation no.", page
225.
224.
Page 107