Number Of Tests; Ok (Result Indicator); Others List; Parameters Menu - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Number of tests

OK (result
indicator)

Others list

Parameters
menu
This is the number of times you wish to repeat the method on the
same sample. The method will be repeated in a new beaker. If the
method is part of a coupled method the number entered here will not
be taken into account. It will be the number of tests entered in the
Coupled method parameters that will be used.
For a reagent calibration method, it is the number of beakers to pre-
pare for the calibration. The average value will be taken into account
when calculating the reagent concentration.
Enter in:
Edit method menu.
Edit reagent menu if Titre = Calibrate.
Range available:
1 to 99 (1 to 20 for a reagent calibration method).
Refer to "Result indicators", page
Choice enabling you to enter electrodes and reagents other than
those from Radiometer Analytical.
For a sample method,
For an electrode calibration method,
see "Electrode calibration parameters", page
For a reagent calibration method,
see "Reagent calibration parameters", page
TitraLab 870 Reference Manual
206.
see "Method parameters menu", page
153.
112.
191.
Page 167

Advertisement

Table of Contents
loading

Table of Contents