Starting up

Set up the instrument in a properly ventilated place. The power supply connector on the rear
panel must remain accessible and close to the user (2 m maximum) so that you can quickly
disconnect the cables in case of emergency.
The room temperature must be between 5 and 40°C.
The relative humidity must be between 20 and 80 %.
To a mains supply socket
Connect the mains socket of the TIM870 to the mains supply using the 3-lead power cord
provided.
The TIM870 must be connected to an earthed mains socket for safety reasons.
Efficient grounding is essential for reliable measurements and security.
In the USA or Canada, use a UL listed power cord only.
Switch on the Titration Manager (O/I power switch set to "I")
The Titration Manager displays during a few seconds an identification screen (name and em-
bedded software version) followed by the Main window,
see "Main window", page
147.