Sample Unit; Sample Unit Conflict; Select Electrode; Select Method - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Sample unit

Sample unit
conflict

Select electrode

Select method

Page 226
Unit of the sample amount. The sample unit can be expressed in
weight (g, mg) or volume (ml, µl) except for the ISE standard addition
method where volume (ml or µl) units are only available.
Enter in:
Edit method > Sample menu
In a Coupled method, the sample unit is not the same for the two titra-
tion methods.
Check the Sample screens of each method of the Coupled method.
Routine user
Is able to select an electrode to check the parameters and/or start a
calibration.
1. Select the method/sequence using the electrode.
2. Press 1 to start a calibration or 3 to check the electrode
parameters. In both cases, the electrodes available will be those
specified in the actual working method or sequence.
Supervisor
From the Electrode window.
1. Press 4.
2. In the ID field, select the electrode from the User list.
To select a method (sample changer not in use):
1. Select Working mode = Method in the Main window.
2. Press 3.
3. In the ID field, select the method from the User list.
To select a method to be run using a Sample Changer:
1. Select Working mode = SAC Method in the Main win-
dow. If this option is not available, enter the Setup menu >
Configuration and define a sample changer.
2. Press 3.
3. In the ID field, select the method from the User list.
TitraLab 870 Reference Manual