Standard Menu; Standard Potential; Standard Solution (Conductivity Measurements); Standard Solution (Ise Measurements) - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Standard menu

Standard
potential
Standard
solution
(conductivity
measurements)
Standard
solution (ISE
measurements)
Standard
solution
(reagent
calibration)
This menu is available for reagent that are edited with the
Titre = Calibrate option. In this menu, identify and enter
the concentration of the standard solution that will be used for the
reagent calibration.
To access:
1. From the Reagent window, press 4.
2. Select the reagent to be edited.
3. Press 2 and check that the Titre = Calibrate option has
been selected.
4. Use the LEFT/RIGHT arrow keys to move to the last display.
5. Press 2 Standard.
Refer to "Direct ISE measurement method - definition", page
Refer to "Standard (conductivity standard)", page
Refer to "Solution menu", page
Refer to "Standard menu", page
TitraLab 870 Reference Manual
235.
233.
237.
95.
Page 237