TitraLab
®

Copy electrode

Page 84
This procedure is used to create an electrode, without having to enter
data nor program the corresponding calibration method.
1. Enter the Electrodes window.
2. Press 4 then 1.
3. In the Function field, select the function according to the elec-
trode type then press ✓,
4. Press ✓ .
5. Select From = Catalogue.
6. In the ID field, select an electrode name from the Catalogue list.
7. In the id field, you can identify the electrode by assigning a sec-
ond id name. This electrode will be called "ID id".
8. Press 1 to confirm then 2 Copy from electrode.
9. In the Library field, select the Preprogrammed or User list.
10. In the ID field, select the electrode to be copied from the list of
available electrodes.
11. Press 1 to confirm. The electrode is created and saved in the
User list.
If you selected the option Preprogrammed, the list is limited to
electrodes of the same type as the "copied" electrode.
If you selected User, the list is limited to electrodes having the
same function (pH, mV (i=0), mV(i>0), ISE, T°C, Reference,
Conductivity or Ground) as the "copied" electrode.
It is not necessary to select Catalogue to create an
electrode using the copy function. An electrode ID created from
Other can also use the copy function.
TitraLab 870 Reference Manual
see "Electrode type", page
118.