TitraLab
®
Assistant func-
tion

Autochaining

Embedded instructions on the Titration Manager display to guide the
user step-by-step through electrode and titrant installations.
These instructions appear at the start of a run method if the working
system has not been correctly installed.
By default this option is set to Yes. It is recommended to use
the default setting at all times!
If the setting is set to No, the Titration Manager considers that the
working system is correctly installed at the start of a run method.
However, this may not be the case, the user must know the status of
the working system at all times!
Enter in:
Setup menu > Configuration menu
This option is valid for a Coupled method which is not part of a
sequence.
Autochaining = No
At the end of each method run, you must confirm the result in order to
perform the next method. If a Notification message has been
selected, a message is displayed between each method of the Cou-
pled method.
Autochaining = Yes
At the end of each method run, The methods are chained
automatically in the Coupled method. If a Notification message has
been selected, a message is displayed upon starting the first method
(no message is displayed after).
Refer to "Notification message", page
Enter in:
Edit method menu (for a Coupled method)
TitraLab 870 Reference Manual
162.