Molar Weight; Monotonic Ip Method; Nat. Water (Temperature Correction) - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Molar weight

Monotonic IP
method
Nat. water
(temperature
correction)
Page 160
When the result unit = weight (g, mg, µg, g/l, mg/l, g/kg, mg/kg,%),
enter the molar weight in g/mol of the species to be determined.
Enter in:
Edit method > Results
Range available:
0.001 to 1000 g/mol.
For this type of method, a preset volume of titrant (increment) is
added to the solution as soon as the measurement variations satisfy a
stability criterion or at the end of a preset acceptation time.
E or pH
Figure 29: Monotonic IP method - size of the increments
In Monotonic IP, one measurement is saved per increment added.
Select a Monotonic IP method when you want to minimise the number
of increments added and in some conditions to speed up a sample
titration.
The Monotonic IP method is stopped after a given number (1 to 8) of
equivalent points has been detected.
Refer to "Temp. correction None/Linear/Nat. water", page
TitraLab 870 Reference Manual
246.