Rinse Burette; Rinse Time - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Rinse burette

Rinse time

For safe reagent handling rinse the burette each time the burette is to
be replaced. It makes sure the delivery tubes and tubings are reagent
free. The rinse cycle lasts for approximately 5 minutes.
To rinse a burette:
1. Enter Burette functions menu:
1. Select the burette in the Address field,
2. Press 4 Rinse.
The rinse procedure runs as follows:
• Empties the burette.
• Prompts you to remove the reagent bottle from the specified
address.
• Fills the burette with air, then empties.
• Prompts you to install the rinse bottle.
• Runs the rinse programme using the rinse solution.
• Prompts you to remove the rinse bottle.
• Fills the burette with air, empties, then refills.
When a Sample Changer is in use enter the time (in minutes and
seconds) the electrodes should be immersed in each static rinse
beaker.
Refer to "Number of static rinses", page
Enter in:
Setup menu > Configuration (if Sample changer = SAC80,
SAC90, SAC850 or SAC950)
Range available:
00:00 to 30:59 min:s
TitraLab 870 Reference Manual
see "Address", page
166.
50.
Page 211