Electrode Calibration Not Required; Electrode Calibration Parameters - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Electrode
calibration not
required
Electrode
calibration
parameters
Page 112
Message appears at the start of a sequence, if a method sequence
has been programmed with an electrode calibration. The electrode
used has been programmed without calibration Calibration
request = No.
Go to Sequence/Sample stack, Edit sequence menu and
remove the electrode/reagent calibration method.
This menu contains the general parameters concerning the
electrode calibration method (measurement stabilisation criteria in
particular).
To access:
1. From the Electrode window, press 4.
2. Select the electrode to be edited.
3. Press 2 Edit electrode and check that the Calibration
request = Yes option has been selected.
4. Edit the electrode calibration general parameters.
5. Use the LEFT/RIGHT arrow keys to move to the last Edit
electrode display.
6. Press 1 Calibration parameters.
TitraLab 870 Reference Manual

Advertisement

Table of Contents
loading

Table of Contents