Electrode Calibration (Ise) - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Electrode
calibration (ISE)
1. Select the method which uses the electrode to be calibrated.
2. Connect the electrode system,
114.
3. Press 1 Calibrate electrodes in the Electrode window.
4. Select the ISE electrode from the list.
5. Press 1 to Run, and follow the messages on the display.
For a 3 to 9-point calibration, the E
S
sensitivity at 25°C and C
25
calculated at the end of the calibration.
For a 2-point calibration, E
For a 1-point calibration, only E
the last multi-point calibration performed or is equal to the default
value (59.16 mV/pC for a monovalent ion). C
The calibration results are saved with the electrode.
To consult the calibration results,
page
134.
It is recommended to maintain all your standards at the same
temperature. Then the temperature entered at the start of a
calibration cycle is valid for all your standards.
TitraLab 870 Reference Manual
see "Electrode connection", page
The TIM870 displays the potential
measured. The displayed temperature is
the temperature measured, entered or is
equal to 25°C according to the
calibration method programmed.
0
standard potential,
detection limit concentration are
0
0
and S
are calculated. C
25
0
is calculated, S
= 0.
0
At the end of a calibration cycle, you can
display the E (mV) = f (pC = -log C)
calibration curve. The calibration points
are marked (here 3). To display the
curve of a calibration cycle, press 2 More
details then 4 Curve from the result
data display.
see "GLP-Archives menu",
= 0.
0
comes from
25
Page 109

Advertisement

Table of Contents
loading

Table of Contents