The Sequence Is Empty; Tim Cell External Gnd; Time Max (Result Indicator); Title - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
The sequence is
empty
TIM cell external
Gnd
Time max
(result
indicator)

Title

Page 248
You are trying to run or complete a sample stack/sequence experi-
ment which contains no programmed methods.
Go to Sequence/Sample stack, Edit sequence screen and check the
sequence.
Specify in the Configuration menu if the grounding of the measuring
cell takes place using an external connection to the titration system.
This is the case when the solution is grounded via a metal shield or
via a conductivity cell connected to a conductivity meter.
The following configurations will be therefore not possible:
• connecting a metal electrode to the GND socket of the titration
system,
• connecting a double platinum electrode to the Pt-Pt socket of the
titration system.
If TIM cell external Gnd = Yes and a reference elec-
trode is connected to the Ref. socket of the titration system,
grounding will take place by an external link and not by the Ref.
socket.
If TIM cell external Gnd = No, grounding of the cell will be
determined in the method by the measurement type (pH, mV or mV
(i>0)) and the parameter Cell grounding (Reference/Metallic/Other).
Refer to "Cell grounding", page
Refer to "Result indicators", page
Enter the title of the calibration report sheet (max. 23 alphanumeric
characters).
Enter in:
Edit method > Printouts
Edit reagent > Printouts
Edit electrode > Printouts
TitraLab 870 Reference Manual