Default Parameters; Delay After Addition; Delete Electrode - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Default
parameters
Delay after
addition

Delete electrode

Page 90
Reset the parameters programmed in the method, reagent, or elec-
trode. Use this command to reset the preprogrammed methods,
reagents or electrodes to the Titration Manager's default values.
Proceed as follows:
1. Display the Main, Reagent or Electrode window.
2. Press 4.
3. Select the method, reagent or electrode ID.
4. Press 3 Default parameters.
5. Press ✓ to confirm the reset.
This delay is counted down after a reagent addition. This time allows
the solution to stabilise after a reagent addition and before the next
titration starts.
While an addition method is running, you can carry on
regardless this delay by pressing key Del.
Enter in:
Edit method > Parameters (titration method if Addition = Yes)
Edit reagent > Calibration parameters (titration method if Addition
= Yes)
Range available:
00:00 to 99:59 min:s
1. Select the electrode to be deleted.
2. Press 4.
3. Press ✓ to confirm or ESC to leave the menu with deleting.
It is not possible to delete an electrode which is used in an-
other method or sequence. Modify the method or sequence,
e.g. change electrode ID or remove the electrode, before
deleting.