Electrode Window; Electrodes Are Different; Empty Burette; Empty Sequence - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Electrode
window
Electrodes are
different

Empty burette

Empty
sequence
This window contains all the information and operations concerning
the electrodes.
To access:
Use LEFT/RIGHT arrow keys.
This message appears if the active electrode used for the reagent
calibration is different from the electrode used in the method.
Check the method and reagent calibration parameters menus.
1. Enter the Burette functions menu.
2. Select the burette in the Address field,
50.
3. Press 2 Empty. The burette is emptied at maximum speed
(approximate emptying time: 25 s).
Involves removing all the methods present in the sequence.
Proceed as follows:
1. Select the sequence to be emptied,
227.
2. Press 2 Sequence/Sample stack then 3 Edit sequence.
3. Press 3 Delete.
4. Press 2 Delete sequence then ✓ to confirm or press Esc to leave
the screen without emptying the sequence.
To remove a particular method from a sequence,
method from a sequence", page
TitraLab 870 Reference Manual
see "Address", page
see "Select sequence", page
see "Remove
197.
Page 119