Sample Preparation No - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Sample
preparation no.
Page 224
The Sample preparation routine gathers all the information required
by a SAC850 or SAC950 Sample Changer to prepare each sample
beakers of the sequence (sample dilution by adding a solvent, stirring
delay, sample degassing, set the sample to a given temperature, start
a dynamic rinse, etc.).
One Sample preparation routine is defined per method executed.
A SAC850 or SAC950 can run up to 10 different sample preparations
in a given sequence.
Sample preparation routines are created and edited from a PC and
TitraMaster 85 PC software. They are loaded in the Titration Manager
memory on connecting the workstation to TitraMaster 85.
On running a sequence, the Titration Manager sends to the sample
changer all sample preparations to be runned for the sequence (one
preparation per method).
Enter in:
Edit sequence menu (with a SAC850 or SAC950)
Range available:
0 to 10
0 means no sample preparation selected.
TitraLab 870 Reference Manual