TitraLab
®
No stirrer (SAC
error)
None
(temperature
correction)
Notification
message
Page 162
SAC950 Sample changer error: there is no magnetic stirring possible
for the beaker position mentionned.
When using a 2 or 3-radii turntable with a SAC950, you must use a
propeller stirrer to stir the solutions that are placed on the smallest
row (2-radii turntable) or on the 2 smallest rows (3-radii turntable).
Green positions : magnetic and propeller stirring possible
Red positions : propeller stirring only
Connect a propeller stirrer to the sample changer. Refer to the Sam-
ple Changer User's Guide, part no. D21T085.
You can also edit a sample stack in order to use a 1-radius turntable
or to use only the largest row of a mutil-radii turntable. In this case, the
magnetic stirring is possible.
Refer to "Temp. correction None/Linear/Nat. water", page
Select Notification = Yes if you want to a message to be
displayed on starting a reagent calibration, a measurement or titration
method. Type the message (3 lines of 1 to 32 alphanumerical charac-
ters).
Enter in:
Edit method menu
Edit reagent menu
TitraLab 870 Reference Manual
Refer to "Sample stack", page
225.