Calibration
stack

Catalogue list

Cell constant
(parameter)
Page 74
For an electrode calibration method,
stack", page
113.
For a reagent calibration method,
page
192.
List of Radiometer Analytical names of electrodes and commonly
used reagents. This list cannot be modified.
Enter the cell constant value. The cell constant is a specification of the
conductivity cell and is normally provided by the cell manufacturer.
If you do not know the cell constant value or if you want to
check its value, select Calibration request =
Yes , edit and run a calibration method.
It is recommended to periodically check the constant value by
performing a cell calibration.
Refer to "Electrode calibration (conductivity)", page
Refer to "Cell constant (definition)", page
Access:
Edit Electrode menu (for a Conductivity type of electrode with
Calibration request = No)
Range available:
-1
0.050 to 15.000 cm
TitraLab 870 Reference Manual
see "Electrode calibration
see "Reagent calibration stack",
(by steps of 0.001 cm
108.
75.
-1
)