Proportional Band; Qc (Result Indicator); Qc Analysis Required - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Proportional
band
QC (result
indicator)
QC analysis
required
Page 182
This is the pH (mV) span prior to the end point, at which the titrant
speed will be reduced. The speed will increasingly reduce as the end
point is approached.
For titrations where the slope of the curve is high near the end point,
the proportional band setting should be large. If overshooting of the
end point frequently occurs, increase the Proportional band
or decrease the Max. speed setting. A high value for the Pro-
portional band will result in slow titrations.
Titrant delivery speed
Maximum
speed
Minimum
speed
Enter in:
Edit method > Parameters menu (end point method)
Edit reagent > Calibration parameters menu (end point method)
Range available:
-9.000 pH to +23.000 pH or -2000.0 mV to +2000.0 mV
Refer to "Result indicators", page
This message is displayed at the start of a method requiring a QC
sample.
Press ✓ and run a QC analysis.
TitraLab 870 Reference Manual
End point no. n
206.