TitraLab
®
Temperature
Probe/ Fixed at
25°C/Entered
Temperature
sensor ID

Test amount

Probe
The sample or standard temperature can be measured using a tem-
perature sensor connected to the Temp socket of the Titration
Manager. Select a temperature sensor in the Method parameters
menu.
Fixed at 25°C
If measurements are to be performed at a constant temperature of
25°C.
Note:
No temperature correction is available for conductivity measurements.
Entered
If the temperature is to be entered manually.
Select in:
Edit method menu
Edit reagent menu if Titre = Calibrate
Edit electrode menu if Calibrate request = Yes
Name of the temperature sensor or name of the electrode with a built-
in temperature sensor.
Enter in:
Edit Electrode > Calibration parameters menu
Edit Method > Method parameters
Edit reagent > Reagent parameters
Range available:
16 characters
Sample quantity to be placed into the beaker.
Enter in:
Menu Beaker (Menu Sequence/Sample stack then press 1 Sample
stack).
Refer to "Sample stack", page
Range available:
0,001 to 100000 (unit = Sample Unit)
TitraLab 870 Reference Manual