Maximum Dose; Measure Type Conflict For A Reprocessing In "Method Id - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Maximum dose

Measure type
conflict for a
reprocessing in
"Method
ID"
Page 150
This parameter is used in conjunction with the Dynamic dose to opti-
mise inflection point determination with respect to the speed of the
titration.
It is the allowed maximum value of each increment to be dispensed by
the burette during an inflection point titration. For titrations where the
inflection on the titration curve is very abrupt (typically redox titra-
tions), use a small value. Decreasing Maximum dose slows down the
titration.
Enter in:
Parameters menu of a Dynamic IP method.
Range available:
0.01 ml to 10 ml.
Manual reprocessing error: The method used for the reprocessing
has a measurement type different from the measurement type of the
curve generated.
Example: you run an End point titration in pH measurement mode and
try to reprocess this curve in manual mode in mV measurement
mode.
Edit the parameters of the method to be reprocessed (parameter
Measurement).
Refer to "Reprocessing a curve (Manual mode)", page
TitraLab 870 Reference Manual
203.