TitraLab
®
Ref. electrode
conflict
Reference
Temp.

Reject a result

Remove burette

Remove method
from a
sequence

Remove reagent

Two reference electrodes are being used in the same beaker.
Change one of the reference electrodes in the electrode system.
Refer to "Temp. correction None/Linear/Nat. water", page
Refer to "Result accepted (Yes/No)", page
Before removing, rinse the burette to remove the reagent.
1. Enter the Burette functions menu:
2. Press 6 Remove burette. The piston is moved to its lowest
position (filling operation).
3. Check the burette data and press 1.
4. Remove the burette and press ✓,
1. Select the sequence to be emptied,
227.
2. Press 2 Sequence/Sample stack then 3 Edit sequence.
3. Select the method to be removed using the LEFT/RIGHT arrow
keys.
4. Press 3 Delete.
5. Press 1 Remove method then ✓ to confirm or press Esc to leave
the screen without removing the method.
The method is removed from the sequence but is not deleted.
To remove all the methods from a sequence,
quence", page
Use this procedure when you no longer wish to use the reagent. To
remove a reagent, run the rinse procedure in the Burette functions
menu.
Refer to "Burette functions menu", page