TitraLab
®

Read me!

An important feature of this Titration Manager interface is that it controls the presence of differ-
ent elements necessary to run the defined application for a selected method/sequence, before
the method/sequence is run.
Working in Supervisor mode
A Supervisor has access to all the libraries for creation purposes.
When programming the Titration Manager in "SUPERVISOR" mode, it is recommended to
work in stages. These stages must be carried out in the order described below:
A.To program method
1. Define your electrode(s)
Identify electrodes (including temperature sensors) to be used for the analysis:
Electrodes can be created from the following lists:
Catalogue,
see "Catalogue list", page
Other,
see "Others list", page
Copy from,
see "Copy electrode", page
When creating the electrode, define if electrode calibration is required (or not), if yes specify
the "periodicity" of the calibrations and the standards to be used.
2. Define reagent
Identify reagents to be used for the analysis
Reagents can be created from the following lists:
Catalogue,
see "Catalogue list", page
Other,
see "Others list", page
Copy from,
see "Copy reagent", page
When creating the reagent, define if reagent calibration is required (or titre entered manually),
if yes specify the "periodicity" of the calibrations and the calibration method.
calibration", page
189.
If a sample changer is to be used, define the sample changer in the Configuration menu before
selecting a SAC sequence.
If you are to perform a calibration, make sure that the electrode(s) used for the
calibration are the same as those used in the method.
Page 18
74.
167.
84.
74.
167.
85.
TitraLab 870 Reference Manual
Refer to "Reagent