Electrode Connection - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Electrode
connection
Page 114
Proceed as follows to connect/install electrodes and temperature
sensors:
1. In the Electrode window, press 2 then 1 Connect electrode.
2. Enter serial number.
3. Install electrodes on the Titration Manager (TIM), ABU52 or
sample changer (SAC80, SAC90, SAC850 or SAC950).
4. Connect electrodes to the rear panel socket of the Titration
Manager (TIM) or ABU52 (ABU1 or ABU2),
page
43. See figure and table below. For example: pHC2001 to
address TIM/E1.
5. Press 1 to confirm.
GND
Socket
REF
Single reference
TEMP
Temperature
GND
Ground metal for cell grounding only
Pt-Pt
Double metal
E1/E2
Indicating (pH, Metal/Redox, ISE) single or combined
EC
Conductivity cell w/o temperature sensor
If the current method in use requires an electrode different to
the one already connected, the Titration Manager will prompt
you to disconnect the electrode before connecting the new
one,
see "Disconnect electrodes", page
Refer to "Electrode connection - Important", page
TitraLab 870 Reference Manual
Refer to "Address", page
EC
t
t
ef
P
-P
R
emp
T
Figure 20: Electrode sockets
Type of electrode
Table 3: Connecting electrodes
see "ABU1/ABU2",
50.
E2
E1
97.
115.