Ec Socket - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

EC socket

6-pin DIN socket for connection of the conductivity cell with 2, 3 or 4
poles and a temperature sensor.
Pin layout:
3
2
4
Pin 1 : pole no.1
Pin 2 : pole no.2
6
Pin 3 : pole no.3
1
5
Pin 4 : pole n°4, also connected to pin no. 5
Pin 5 : 0 V (ground)
Pin 6 : temperature sensor
Potential imposed between poles 2 and 3: ±200 mV constant for all
conductance ranges.
The current passing through poles 1 and 4 is measured. The potential
between poles 1 and 4 cannot exceed 3 V in absolute value.
The following Radiometer Analytical conductivity cells can be
connected to the EC socket:
Conductivity
Number
cell
of poles
CDC566T (*)
CDC866T (*)
CDC641T (*)
CDC511T
CDC741T (*)
CDC861T
CDC565
CDC749
CDC745-9 (*)
CDC267-9
CDC241-9
XE100
Figure 19: Radiometer Analytical conductivity cells
(*)
This conductivity cell is present in the TIM870 electrode library (Catalogue list)
Conductivity cell with 2, 3 or 4 poles?
see the "Conductivity theory and practice" guide, part no. D61M002.
TitraLab 870 Reference Manual
Built-in
temperature
sensor
4
Yes
4
Yes
2
Yes
4
Yes
2
Yes
4
Yes
4
No
2
No
2
No
2
No
2
No
2
No
Connection to
TIM870
Direct connection
Direct connection
Direct connection
Direct connection
Direct connection
Direct connection
Direct connection
Direct connection
Via A94L136 cable
Via A94L136 cable
Via A94L136 cable
Via A94L136 cable
Page 103