Sample Stack; Sample Type Incompatible - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Sample stack

Sample type
incompatible
The sample stack gathers individual data for all the samples present
in a sequence.
1. Perform steps 1 to 8 of the "Programming sequence" procedure.
Refer to "Programming sequence", page
2. In the Main window, press 2 Sequence/Sample stack.
3. If required enter a new sequence ID.
4. Press 1 to enter the sample stack.
5. Enter sample data.
Label the beakers indicating the number of beakers in the se-
quence, for example: 1/11, 2/11 etc.... the sample ID and the
number of times the method is to be performed in the beaker.
Place the beakers in the numbered position on the sample
changer. If static rinses and/or dynamic rinses are
programmed, position the corresponding rinse beakers at the
right places.
Refer to "Number of static rinses", page
Refer to "Dynamic rinses", page
You can print the sample stack by pressing Print from the
Sequence/Sample stack screen (screen displayed at step 2
of the operating procedure above).
The method used for the reprocessing is a sample method and the
curve is a calibration curve or reverse.
Select the correct sample type method.
TitraLab 870 Reference Manual
181.
<1/11> means the first beaker over 11
programmed in the sequence.
Use the LEFT/RIGHT arrow keys to re-
view the other beakers in the sequence.
The data displayed will depend on the
type of method selected.
166.
102.
Page 225