The reagent calibration stack defines individual data for each
standard solution beakers present in a reagent calibration sequence.
1. In the Configuration menu, declare the model of sample changer
used (SAC80, SAC90, SAC850 or SAC950).
Refer to "Configuration menu", page
Depending on the model of sample changer used, enter the
sample changer configuration parameters.
Refer to "Sample changer", page
2. In the Main window, select SAC Sequence or SAC method
for the working mode. This sequence or method must use
the reagent you want to calibrate. Edit the sequence or the
method if relevant,
3. Enter the Reagent window and press 1 Calibrate/Enter titre.
4. Select the reagent to calibrate.
5. Press 3 Calibration sequence.
Label the beakers indicating the running number in the
sequence, for example: 1/5, 2/5 etc.... and the name of the
standard solution.
Place the beakers in the numbered position on the sample
changer.
If rinses are programmed, position the corresponding rinse
beakers at the right places.
Refer to "Number of static rinses", page
Refer to "Number of dynamic rinses", page
You can print the calibration stack by pressing Print from the
calibration menu.
6. Press Esc then run the sequence by pressing 1 Run calibration.
TitraLab 870 Reference Manual
81.
221.
see "Programming sequence", page
<1/5> means the first beaker over 5
programmed in the sequence. Use the LEFT/
RIGHT arrows to review the other beakers in
the sequence.
The Standard name and sample amount are
entered in the Standard menu of the reagent
(ID and Standard amount parameters).
Enter the batch number of each standard
solution.
181.
166.
164.