Running A Method; Running A Reagent Calibration Sequence - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Running a
method
Running a
reagent
calibration
sequence
Page 216
Before running a method, perform the following operations in the
order listed below:
1. Select method.
Refer to "Select method", page
2. If a Question mark "?" is present in the Reagent and/or Electrode
tabs, it means that the method needs to be programmed - a rea-
gent or an electrode is missing. Review programming in Supervi-
sor mode.
Refer to "Programming method", page
If you have a problem running the analysis, the Titration
Manager will guide you through the necessary operations so
that you are able to run the analysis in no time at all.
3. Connect/calibrate the electrode(s).
Refer to "Electrode connection", page
Refer to "Calibrate electrodes", page
4. Install/calibrate reagent(s).
Refer to "Install reagent", page
Refer to "Calibrate reagents", page
5. When Sunny icons are visible in the Reagent tab, press 1 in the
Main window to run the method.
Refer to "Reagent icons", page
6. If prompted to do so, enter the User's name (ID) and press 1.
7. Enter the Sample name (ID) and press 1.
8. Dip the electrodes into the sample beaker and press 1 to
continue.
For the following steps:
Refer to "Run window", page
Refer to "Reagent calibration (SAC sequence)", page
TitraLab 870 Reference Manual
226.
114.
72.
140.
72.
193.
213.
180.
190.