Testing Readiness for Operation

Testing Readiness for Operation
Before using for the first time
1 Check that the mains voltage matches the values indicated
on the rating plate next to the power cable.
2 Plug into the mains.

Before each use

Check that the unit has been disinfected.
Check that the side panels are locked securely into place.
The panel hardware must be visible above the bed surface.
Check that the side panels are free from cracks and sharp
edges.
Check that the correct mattress is in the cot;
Operation with mattress heater: gel mattress
Operation without mattress heater: foam mattress.
Check that the bed tilts properly and locks securely into
position.
Check that the required accessories and therapy
equipment are available and in proper working order.
Check that the gas supply is available and sufficient for the
accessories and equipment to be used.
Check that the cables and hoses are correctly and securely
installed. Never route cables or hoses over the panels
because they might be pinched or crushed when folding up
the panels or fitting the bed canopy.
Checking height adjustment (optional)
Test height adjustment system if fitted:
3 Press the right pedal briefly to raise the bed.
4 Press the left pedal briefly to lower the bed.
Remember the maximum load for height adjustment
is 50 kg.
Hoses and cables should be long enough to ensure
a secure connection even in the top or bottom height
adjustment position.
Do not place any objects in the raising/lowering path.
Adjust to a comfortable working height.